EX36x‐en‐GB_V1.61/17
5
SafetySymbolsthataretypicallymarkedonmetersandinstructions
Thissymbol,adjacenttoanothersymbol,indicatestheusermustrefertothe
manualoruserguideforfurtherinformation.
Riskofelectricalshock
Fusesymbol
Equipmentprotectedbydoubleorreinforcedinsulation
Batterysymbol
ConformstoEUdirectives
Donotdiscardthisproductinhouseholdtrash.
ACmeasurement
DCmeasurement
Earthground
UnsafeVoltageAlert
Whenthemeterdetectsavoltageequaltoorgreaterthan30Voravoltageoverload(OL)inV,mV,orLoZ
mode,thesymbol
isdisplayed.Thissystemwasdesignedtoalerttheuserofapotentiallyhazardous
voltage.
PERIEC1010OVERVOLTAGEINSTALLATIONCATEGORY
OVERVOLTAGECATEGORYI
EquipmentofOVERVOLTAGECATEGORYIisequipmentforconnectiontocircuitsinwhichmeasuresare
takentolimitthetransientover‐voltagestoanappropriatelowlevel.
Note–Examplesincludeprotectedelectroniccircuits.
OVERVOLTAGECATEGORYII
EquipmentofOVERVOLTAGECATEGORYIIisenergy‐consumingequipmentto
besuppliedfromthefixed
installation.
Note–Examplesincludehousehold,office,andlaboratoryappliances.
OVERVOLTAGECATEGORYIII
EquipmentofOVERVOLTAGECATEGORYIIIisequipmentinfixedinstallations.
Note–Examplesincludeswitchesinthefixedinstallationandsomeequipmentforindustrialusewith
permanentconnectiontothefixedinstallation.
OVERVOLTAGECATEGORYIV
EquipmentofOVERVOLTAGECATEGORYIVisforuseattheoriginoftheinstallation.
Note–Examplesincludeelectricitymetersandprimaryover‐currentprotectionequipment