User Manual
EX330A-en-GB_V5.4 10/17
13
Resistance
400
0.1
(1.2% reading + 4 digits)
4k
1
(1.2% reading + 2 digits)
40k
0.01k
400k
0.1k
4M
0.001M
40M
0.01M
(2.0% reading + 3 digits)
Capacitance
4nF
0.001nF
(3.5% reading + 40 digits)
40nF
0.01nF
400nF
0.1nF
(2.5% reading + 4 digits)
4F
0.001F
(3.5% reading + 4 digits)
40F
0.01F
200F
0.1F
(3.5% reading + 10 digits)
Frequency
5.000Hz
0.001Hz
(0.1% reading + 1 digits)
50.00Hz
0.01Hz
500.0 Hz
0.1Hz
5.000kHz
0.001kHz
50.00kHz
0.01kHz
500.0kHz
0.001MHz
5.00MHz
0.01MHz
10.00MHz
0.01MHz
Sensitivity: 0.8V rms min. @ 20% to 80% duty cycle and <100kHz; 5Vrms min @ 20% to 80% duty
cycle and > 100kHz.
Duty Cycle
0.1 to 99.9%
0.1%
(1.2% reading + 2 digits)
For duty cycle the pulse width range is 100µs - 100ms (Frequency: 5Hz to 150kHz)
Temp (type-K)
-4 to 1382F
1F
(3.0% reading + 8 digits)
(probe accuracy not included)
-20 to 750C
1C
NOTES:
Accuracy specifications consist of two elements:
(% reading) – This is the accuracy of the measurement circuit.
(+ digits) – This is the accuracy of the analog to digital converter.
Accuracy is stated at 65
o
F to 83
o
F (18
o
C to 28
o
C) and less than 75% RH.