Installation Guide

DO600-en-GB_V7.4 12/16
4
Calibration
1. Calibrationshouldbeperformedonadailybasis.
2. TurnthemeterON
3. PressandholdtheMODE/HOLDbuttonuntil%isdisplayedontheLCD.
4. Ifthemeterhasnotbeenusedforseven(7)daysorlonger,allowtheelectrodetofully
polarize.Thismaytake23minutes.
5. Placetheelectrodecapontotheelectrode.Thespongecontainedinthecapshouldbeonly
moistened(notsoaked)with(distilledwater)orcleantapwater.Besurethattheelectrode
membraneiscleananddryorthecalibrationwillbeincorrect.Thereneedstobeanairgap
betweenthemembraneandthesponge.Nevertouchthemembrane,asskinoilwillaffect
theelectroderesponse.
6. WaituntilthereadingstabilizesthenpressandholdtheCAL/RECALLbuttonuntilCALis
showninthelowerdisplay.Thereadingswillblink“101.7”and“SAwillappear.
7. Whenthecalibrationiscomplete“End”willappearandthenthemeterwillreturntothe
normalmeasurementmode.Turnoffthemeter.
Note:The“SA”willnotappearifthecalibrationfails.
Optional’zerooxygen’calibration(improvesmeasurementaccuracyforveryloworveryhighDO
measurements):
Placetheelectrodeinazerooxygencalibrationsolution,suchas5%sodiumsulfite,waitfor
stabilityandpresstheCAL/RECALLbuttonuntilCALisshownintheinthelowerdisplay.Stabilityin
azerosolutionmaytakemanyminutes,dependingonelectrodehistory.
Note:SodiumSulfitecanbecomedepositedontheelectrodeandonthe“coined”surfaceofthe
electroderetainingcollar.ThepresenceoftheSodiumSulfitewillnegativelyaffectfutureDO
measurementsuntilsuchatimewhereALLofitisremovedfromtheelectrode.
OptionalElectronicZerocalibration.
IftheDO600isnotmakingstablemeasurementsoryouarereplacingtheelectrodewithanew
assembly,performthisZerocalibration.
Turnoffthemeter,removetheelectrodebyunscrewingtheretainingringandpulloffthe
electrode.
TurnonthemeterandwaitfortheSelfcalibrationtocomplete.

SettheModeto%.
PressandHoldtheCALbutton untilCALappearsonthedisplay.Whenthecalcyclecompletes,the
displayshouldread0.0%.
Turnoffthemeter.
Reattachtheelectrode.PerformtheelectrodecapcalibrationfromStep5.