CALIBRATION
TheCalibrationmenuallowstheusertoperformaFerrous(ZeroFE)ornonferrous(ZeroNFE)
ZEROCalibration.TheCalibrationmenualsoallowstheusertodeleteZEROcalibrationdata for
bothferrous(ZeroF)andnonferrous(ZeroN)modes.
1. PresstheLEFTbutton(SEL)toaccessthe
mainmenu.
2. UsetheDOWNbuttontoscrolltoCALIBRATION.
3. PressLEFT(SELECT)toopentheCALIBRATIONparameter.
4. ToperformaZEROCalibration,refertoZeroCalibrationinthenextsection.
5. TodeleteZEROCalibrationdata,scrolltoDELZeroofFEorDELZeroof
NFEandthenpress
theLEFTbutton(SELECT).ThisexecutestheZerodeletefunctionandreturnsthemeterto
theSETmenu.
6. PresstheRIGHTbutton( ) toreturntothenormaloperatingmode.
ReferenceFilmAccuracyVerification
1. TheCG206metercomeswithasetofreferencefilmstocheckthemeasurement
accuracyofthemeter.
2. ThebasematerialsareFerrous(Steel)andtheNonFerrous(Aluminum).
3. Turnonthemeter1meterawayfromanymaterial.
4. SettheMeasureModetoFE,NFE,or
Auto.
5. PerformaZerocalibrationonthebasematerialofyourchoice.
6. Placeathicknessfilmonthebaseandmakeameasurement.
7. Refertotheaccuracyspecifications tovalidatetheaccuracyofthemeter.