User Manual
CG206‐en‐GB_v1.06/16
8
CALIBRATION
TheCalibrationmenuallowstheusertoperformaFerrous(ZeroFE)ornon‐ferrous(ZeroNFE)
ZEROCalibration.TheCalibrationmenualsoallowstheusertodeleteZEROcalibrationdata for
bothferrous(ZeroF)andnon‐ferrous(ZeroN)modes.
1.  PresstheLEFTbutton(SEL)toaccessthe
mainmenu.
2.  UsetheDOWNbuttontoscrolltoCALIBRATION.
3.  PressLEFT(SELECT)toopentheCALIBRATIONparameter.
4.  ToperformaZEROCalibration,refertoZeroCalibrationinthenextsection.
5.  TodeleteZEROCalibrationdata,scrolltoDELZeroofFEorDELZeroof
NFEandthenpress
theLEFTbutton(SELECT).ThisexecutestheZerodeletefunctionandreturnsthemeterto
theSETmenu.
6.  PresstheRIGHTbutton( ) toreturntothenormaloperatingmode.
ReferenceFilmAccuracyVerification
1.  TheCG206metercomeswithasetofreferencefilmstocheckthemeasurement
accuracyofthemeter.
2.  ThebasematerialsareFerrous(Steel)andtheNon‐Ferrous(Aluminum).
3.  Turnonthemeter1meterawayfromanymaterial.
4.  SettheMeasureModetoFE,NFE,or
Auto.
5.  PerformaZerocalibrationonthebasematerialofyourchoice.
6.  Placeathicknessfilmonthebaseandmakeameasurement.
7.  Refertotheaccuracyspecifications tovalidatetheaccuracyofthemeter.