User Manual

CG206enGB_v1.06/16
2
Introduction
ThankyouforselectingtheExtechCG206CoatingThicknessTester.TheCG206isaportable
meterdesignedfornoninvasivecoatingthicknessmeasurementswithautomaticrecognitionof
thesubstratematerialundertest.
Themeterusestwomeasurementmethods:magneticinduction(forferrousmetalsubstrates)
andeddycurrent(fornonferrousmetal
substrates).
TheCG206metercanconnecttoaPCviaUSBandtransmitrecordeddataforfurtheranalysisand
forgeneratingreportdocumentation.
Properuseandcareofthismeterwillprovidemanyyearsofreliableservice
.
Features
MeasuredCoatings:Nonmagneticcoatings(e.g.paint,zinc)onsteel;InsulatedCoatings(e.g.
paint,anodizedcoatings)onnonferrousmetals
Intuitivemenubasedprogramming
Memorystorageofupto1500readingsin30groups(50readingspergroup)
Deletesinglereadingsorallgroupseasily
Auto
PowerOFFwithdisablefeature
DatatransfertoPCviaUSB
UserZeroCalibrationability
Applications
Corrosionprotection
PaintShopsandElectroplaters
Chemical,Automobile,Shipbuilding,andAircraftindustries
Laboratory,Workshop,andFielduse
ProbeDescription
TheCG206probeisspringmountedinthesensorsleeve.Thisensuressafeandstablepositioning
oftheprobeandensuresconstantcontactpressure.
AVgrooveinthesleeveoftheprobefacilitatesreliablereadingsonsmallcylindricalparts.The
hemisphericaltipoftheprobeismadeofhardand
durablematerial.
Measuringmetalliccoatings
Thismetercanmeasurenonmagneticmetalcoatings(Zinc)onamagnetic(ferrous)substance
base,andnonmetalcoatingsonametalbase(ferrousornonferrous).