User`s guide
EX430-EU-EN v5.5 08/13
4
OVERVOLTAGE CATEGORY III
This meter meets the IEC 610-1-2001 standard for OVERVOLTAGE CATEGORY III. Cat III meters
are protected against overvoltage transients in fixed installation at the distribution level. Examples
include switches in the fixed installation and some equipment for industrial use with permanent
connection to the fixed installation.
SAFETY INSTRUCTIONS
This meter has been designed for safe use, but must be operated with caution. The rules listed
below must be carefully followed for safe operation.
1. NEVER apply voltage or current to the meter that exceeds the specified maximum:
Input Protection Limits
Function Maximum Input
V DC or V AC 1000V DC/750V AC, 200Vrms on 400mV range
mA AC/DC 500mA 250V fast acting fuse
A AC/DC 20A 250V fast acting fuse(30 seconds max every 15
minutes)
Frequency, Resistance, Capacitance,
Duty Cycle, Diode Test, Continuity
250Vrms for 15sec max
Temperature 60V DC/24V AC
2. USE EXTREME CAUTION when working with high voltages.
3. DO NOT measure voltage if the voltage on the "COM" input jack exceeds 600V above earth
ground.
4. NEVER connect the meter leads across a voltage source while the function switch is in the
current, resistance, or diode mode. Doing so can damage the meter.
5. ALWAYS discharge filter capacitors in power supplies and disconnect the power when making
resistance or diode tests.
6. ALWAYS turn off the power and disconnect the test leads before opening the covers to replace
the fuse or battery.
7. NEVER operate the meter unless the back cover and the battery and fuse covers are in place
and fastened securely.