Specifications

382357EUENV2.42/13
5
HighandLowAlarm
1. Turnthemeteronbymovingtherotaryswitchtothe)))position.
2. PresstheFUNCbuttontoviewandadjustthe“HI”alarmvalue.
3. Pressorthebuttontoincreaseordecreasethevalue.Holddownthebuttonto
increasethespeedofadjustment.Thevaluecanbeincrementedfrom0to1500ohmsand
thenOL(overload).Thevaluewillrolloverto0afterOLisdisplayed.
4. Oncethedesiredvalueisset,presstheFUNCbuttona secondtimetoviewandadjustthe
“LO”alarmvalue.
5. PresstheFUNCbuttonthreemoretimestoexitthesettingmode.
6. Themeterwillnowcomparethemeasuredresistancevaluewiththehighandlowlimits
whenthemeterissettothe)))position.IfthereadingislargerthantheHIvalue,the
meterwillbeepanddisplayHI”ontheLCD.Ifthereadingislowerthan theLOvalue,the
meterwillbeepanddisplayLO”ontheLCD.
DataLogging
SettingtheSamplingTime
1. PresstheFUNCbuttonthreetimesuntiltheSECsymbolappearsintheLCDdisplay.
2. Thesamplingrateisnowdisplayedinseconds.
3. Presstheorthebuttontoincreaseordecreasethevalue.Thevaluecanbe
incrementedfrom0to255seconds.
4. Oncethedesiredvalueisset,presstheFUNCbuttonseveraltimesuntilthecharactersin
theupperrowoftheLCDdisappear.
RecordingData
1. PresstheRECbuttontostartrecordingdataatthespecifiedsamplingrate.TheREC
symbolwillappearonthetopoftheLCDdisplay.Tostoprecording,presstheRECbutton.
TheRECsymbolwillnolongerbedisplayedontheLCD.
NOTE:Dataloggingwillautomaticallystop
ifthememoryisfilled(116records),orifthemeter
detectsalowbatterycondition.
NOTE:Ifthesamplingintervalissetto0,onlyonesamplewillberecorded.Torecordanother
sample,pressRECagain.