RS-232 PC Communications Interface
RS-232 data format
Baud rate 9600
Parity none
Data bits 8
Stop bits 1
Flow control none
RS-232 Connection
Connect the supplied communications cable to the rear of the power
supply and to the communications port of the PC.
Power Supply PC
Pin 2 ------------------------------ Pin 2 RX
Pin 3 ------------------------------ Pin 3 TX
Pin 4 ------------------------------ Pin 4 DTR
Pin 5 ------------------------------ Pin 5 GND
RS-232 Commands
The command format is as follows: Command
Parameter <Carriage Return (cr)>
For example: V 20.5 (sets output to 20.5VDC)
An ASCII code for <Carriage Return> must be sent together with the command or commands.
Commands will not be processed until a <Carriage Return> is received. Note that the unit can
process up to 50 characters. If more than 50 characters are received, the unit will clear the entire
command buffer.
Summary of Commands
STBY <cr> (place the unit in standby mode)
OPER <cr> (place the unit in operating mode)
V xx.xxx <cr> (set the voltage output)
A x.xxx <cr> (set the current output)
SEC xxx.xx <cr> (enter time in seconds)
OVP xx.xxx <cr> (enter desired voltage value for over-voltage protection)
OCP 0 <cr> (disable the over-current protection)
OCP 1 <cr> (enable the over-current protection)
STORE xxx <cr> (store the present output values to memory location xxx)
RECALL xxx <cr> (recall an output setup from memory location xxx)
STEP <cr> (enable the ATS Auto Test Sequencing function)
? <cr> (retrieve all data)
V? <cr> (retrieve the voltage value)
A? <cr> (retrieve the current value)
OCP? <cr> (retrieve the status of OCP)
OVP? <cr> (retrieve the voltage value for OVP)
SEC? <cr> (retrieve the time value in seconds)
STATUS? <cr> (retrieve the STBY/OPER, OCP, OVP, & 5V/3.3V fixed output status)
Note: All commands are case insensitive.