Specifications

382275/6-EU-EN v4.3 3/14
10
RemoteControlMethodB:Usingtwoexternalvariable5kohmresistors
1. EnsurethattheloadisdisconnectedandthatthepowersupplyisOFF.
2. Preparetwo5kohmvariableresistorsasshowninthediagrambelowandusethe
wiresfromPort1,2and4asshown.
3. SwitchthepowersupplyON.
4. AdjustthePin2
5kohmvariableresistorfromoneendtootherendtocheckand
verifythefulloutputvoltagerangeofthepowersupply.
5. Shortthemainoutputwith10AWGwireandchecktheCCdisplaybyadjustingthe
Pin35kohmvariableresistor.
6. SwitchOFFthepowersupply.

RemoteSocketPINassignmentfor externalvariableresistors
PIN FUNCTION NOTES
1 Internal+5VDC
2 VoltageAdjustment Variablepartofresistor
3 CurrentAdjustment Variablepartofresistor
4 Ground
5 OutputOFF Shorttoground
6 Unused
7 Unused
8 Unused
