Electromagnetic Compatibility Compliance Statement Information
The Willow Breast Pump Generation 3 may be used at home, office, or other general areas.
Special consideration should be given to use:
During Medical Procedures: Do not use the Willow pump during any medical
procedure or during imaging. For example, do not use
pump during RF ablation of a skin anomaly, x-ray, MRI, or ultrasound.
With other Equipment: Use of this equipment adjacent to or stacked with other
equipment should be avoided because it could result in improper
operation. If such use is necessary, the pump and the other equipment
should be observed to verify that they are operating normally.
In High Electrostatic Discharge Environments: If the air is very dry, use the pump
where oors are wood, concrete, or ceramic tile. If oors are covered with
synthetic material, the relative humidity should be at least 30%. If you notice
shocks when you touch things, as sometimes happens in very dry winter months,
be sure to touch a metal item such as a table before handling the Willow pump.
In Industrial Settings: Avoid using the pump near arc welding equipment,
radar installations, and similar industrial settings.
Near RF Emitters: Keep devices that use WiFi, Bluetooth, and other intentional
RF emitters at least 30 cm from the pump during operation. This includes
communications equipment peripherals such as antenna cables and external
antennas. Otherwise, degradation of the performance of the Willow pump
could result. If you know the power output the RF emitter, the table labeled
“Electromagnetic Compatibility Compliance Statement Information”
can be used for more precise information about separation distances.
TECHNICAL SPECIFICATIONS
POWER REQUIREMENTS
Only use the Willow Breast Pump 1, 2, or 3 Charger
Manufacturer: Delta Electronics, Inc.
Part Number: MDS-030AAC05 ZB | Specications: 100 - 240 Vac, 50-60Hz, 0.5A-0.3A
Part Number: MEF-010A05B JA | Specications: 100 - 240 Vac, 50-60Hz, 0.4A-0.2A
SPECIFICATIONS
Pump Dimensions 141mm x 85mm x 112mm
Pump Weight 355 g
Vacuum Strength 35-245 mmHg
Cycle Speed 42-90 cycles/minute
Motor Life
273 hours (Dened to be approximately three 15 minute
sessions per day for one year.)
Bluetooth Info Bluetooth v4.1 (Bluetooth Low Energy); Range up to 10 m
Security Encryption 128-bit AES (Advanced Encryption Standard)
Date of Manufacture: The date of
manufacture can be located within
the serial number of the pump. The
serial number is an 11 digit number
formatted as: xxxYM xxxxxx where
“Y” is the year and “M” is the month.
Use the tables to determine the year
and month of manufacture.
Date of Manufacture:
Year Code
2016
2017
2018
2019
2020
2021
2022
2023
A B C D E F G H
Month Code
Jan Feb Mar Apr May Jun Jul Aug Sep Oct Nov Dec
1 2 3 4 5 6 7 8 9 A B C