User's Manual

EX7100 Technical Manual - Module Descriptions
©1999 Exicom Technologies (1996) Ltd 7-29 Issue 3 - May 2001
7.6.2 Testing and Fault Diagnosis
Note: Refer to Section 5 “Troubleshooting the EX7100” before reading this Section.
The module contains semi-conductor components that are
sensitive to electrostatic damage.
Electrostatic sensitive devices should only be stored and
transported inside electrically conductive static shielding
bags. Fault finding and module replacement must be
performed at “electrostatically safe work stations” where the
work bench surface, soldering iron and the operator are all
earthed to prevent the build up of harmful electrostatic
charges.
The RF Power Amplifier module must not be run without a
50 load on its output. Failure to meet this condition may
damage the amplifier.
The RF devices used in the Power Amplifier module may
contain beryllium dioxide. These devices must not be broken,
ground, hammered or tampered with in any way, and should
be disposed of in a safe manner as specified in local
environmental and safety legislation. If safe disposal is a
problem, they should be sealed in an appropriate container
and returned to Exicom for disposal. In this case, it is the
senders’ responsibility to ensure that all precautions are
taken to ensure safe delivery to Exicom.
UHF RF Power Amplifier Module Testing
Module DC Input Power Measurement
1. Remove the bottom cover of the terminal as described in Section 6.1.
2. Measure the power supply to the RF Power Amplifier module on the Tx PCB with a
multimeter (by directly accessing the pins within PL103). Refer to Table 3.3 below for
expected voltages on PL104.
Pin Number on PL103 Expected Voltage
1 Ground
2 +14 to 16 V
3 +14 to 16 V
4 +22.8 to 25.4 V
Table 7.3: Power Amplifier Module DC Input Voltages