User's Manual

EX7100 Technical Manual - Module Descriptions
©1999 Exicom Technologies (1996) Ltd 7-58 Issue 3 - May 2001
7.11.2 Testing and Fault Diagnosis
The module contains semi-conductor components that are
sensitive to electrostatic damage.
Electrostatic sensitive devices should only be stored and
transported inside electrically conductive static shielding
bags. Fault finding and module replacement must be
performed at “electrostatically safe work stations” where the
work bench surface; soldering iron and the operator are all
earthed to prevent the build up of harmful electrostatic
charges.
The Modem/Controller module contains a lithium battery.
There is a danger of explosion if the battery is incorrectly
replaced. Replace only with the same or equivalent type
recommended by the manufacturer.
Dispose of used batteries according to the manufacturers
instructions.
General
The Modem/Controller module is very complex and interacts with all of the other modules
and control systems within the EX7100 terminal.
For this reason, a faulty Modem/Controller module may cause symptoms that appear
elsewhere, and faults elsewhere may cause symptoms to appear at the
Modem/Controller module.
Accessing the Modem/Controller
To gain service access to the Modem/Controller module, remove the top cover as per
Section 6.1.
Diagnosis
Under normal operating conditions and with a link established, D8, D100, D101 and D102
should all be on. If this is not the case, follow the instructions below:
With the unit de-powered, check the Modem/Controller module visually for signs of
damage such as burnt tracks or components. Check that all connectors are fully mated
and that all socketed ICs are fully inserted.
If no damage is apparent, attach the antenna and power leads. Follow the steps in Figure
7.30 overleaf.