Instruction Manual

P/N: 1051923 January 2008 1/4
ADDENDUM
This information applies only to the
PM-1100 Optical Power Meter user guide.
GPIB Commands and Queries
³ The FORM:READ:DATA? query does not exist.
³ The MMEM:ACQ:DATA:RECA:VALUE? query should be defined as
follows:
MMEMory:ACQuisition:DATA:RECAll:VALUE?
Description This query returns the measurement saved in the <data>
memory location.
Syntax MMEM:ACQ:DATA:RECA:VALUE?<space><data>
Parameters The <data> parameter represents the memory location where
measurement data is stored (out of the 1024 available).
Response Measurement data saved in the specified memory location in
the “±999.999E±99” format.
Notes
³ Use the MMEM:ACQ:DATA:RECA:UNIT? query to get the
corresponding measurement units.
³ Use the MMEM:ACQ:DATA:RECA:WAVE? query to get the
wavelength at which the measurement was taken.
Example MMEM:ACQ:DATA:RECA:VALUE? 1021