Using the Wavelength Calibration Software
DWDM Passive Component Test System 145
Using Calibration Tools
Using Calibration Tools
There are two ways to access this step: 
³ By clicking Tools during the first step.
³ By carrying out the entire process.
³ Apply New: writes the new calibration constants found for the 
IQS-9401 WRM module and displays a calibration certificate. For more 
information on the calibration certificate, see Viewing and Printing a 
Calibration Certificate on page 146. However, if the whole process is 
not performed, you will not be able to access the Apply New button 
because no calibration constants have been found. 
³ Apply Old: Accessible even if the whole calibration process is not 
performed. It re-writes calibration constants, by applying an old 
calibration file or the Factory Default. 
³ Save Current: Accessible even if the whole calibration process is not 
performed. It allows you to save the current calibration constants to a 
binary file identified with the following format name: 
SerialNumber-Date-Hour.dat.