TDR Test
200 FTB-610
Xtalk TDR
 Trace Setup button opens a new page used for trace settings. See 
Trace Setup on page 194 for more information.
 Save/Load Trace(s) button is used for loading and saving traces. See 
Save/Load Trace(s) on page 196 for more information.
 Save Test button opens a new page that allows you to save a snapshot 
of test results into a result file. Refer Save Test on page 74 for more 
information.
 Trace allows you to select corresponding configurations showing the 
following Trace control values: 
 Values: Trace 1 (Live)
 Trace 2 (Static)
 Trace 3 (Static)
 Drop-down lists of the following TDR values allow you to change their 
respective parameters for live Trace 1. Modification of the on-screen 
parameters will be applied to the live trace for the next cycle of 
measurement. When Trace 2/3 is selected, these parameters are 
disabled and the list-values are shown as read only.
 Range
 Gain
 Pulse Width 
 VoP allows you to change the velocity of propagation value for the 
cable using the virtual keyboard. Or to change the value by 0.01, 
use the on-screen up/down arrows or up/down arrow keys of the 
keypad.
Note: Changes to the VoP only apply during the test, for both factory and user 
cables. Upon exiting the test, the original cable VoP value set in the Cable 
Book remains unchanged.