Owner's manual
SCPI Command Reference
548 FTB-5240S/S-P/BP
Product-Specific Commands—Description
DEFault allows the instrument to select a value 
for the <Offset> parameter.
The <Offset> parameter corresponds to a valid 
power offset value.
The 
SENSe[1..n]:CORRection:OFFSet[:MAGNitude]? 
MIN and 
SENSe[1..n]:CORRection:OFFSet[:MAGNitude]? 
MAX queries can be used to determine valid 
power offset range.
Example(s) SENS:CORR:OFFS:MAGN 0.5 DB
UNIT:RAT DB
SENS:CORR:OFFS:MAGN? Returns 
5.000000E-001 
See Also :SENSe[1..n]:WAVelength:OFFSet
:SENSe[1..n]:CORRection:OFFSet[:MAGNitude]?
:SENSe[1..n]:CORRection:OFFSet
[:MAGNitude]