User Manual
Referencia de instrucciones SCPI
OSA 587
Product-Specific Commands—Description
:SENSe[1..n][:WAVelength]:STOP
Description This command sets instrument sweep stop 
wavelength.
At *RST, this value is set to 1570.0 nm.
Syntax :SENSe[1..n][:WAVelength]:STOP<wsp><Stop[
<wsp>M]>|MAXimum|MINimum|DEFault
Parameter(s) Stop: 
The program data syntax for <Stop> is defined 
as a <numeric_value> element followed by an 
optional <SUFFIX PROGRAM DATA> element. 
The allowed <SUFFIX PROGRAM DATA> 
element is M. The <Stop> special forms 
MINimum, MAXimum and DEFault are accepted 
on input.
MINimum allows to set the instrument to the 
smallest supported value.
MAXimum allows to set the instrument to the 
greatest supported value.
DEFault allows the instrument to select a value 
for the <Stop> parameter.
The <Stop> parameter corresponds to a valid 
sweep stop wavelength value.
The SENSe[1..n][:WAVelength]:STOP? MIN and 
SENSe[1..n][:WAVelength]:STOP? MAX queries 
can be used to determine valid sweep stop 
wavelength range.