User Manual
CableSHARK P3 User Guide 
137
MAIN
MENU
EF
CHG
SAVE TO
MEM
SAVE TO
USB
UPLOAD
RESULT
DMMNEXT
TEST
(T - R) AC RMS Voltage
(T - G)
(R - G)
 AC RMS Voltage
 AC RMS Voltage
=
=
=
  0 mV  0 Hz
  0 mV  0 Hz
  0 mV  0 Hz
DC Voltage
DC Voltage
DC Voltage
=
=
=
   0 mV
   0 mV
   0 mV
(T - R) Resistance
(T - G)
(R - G)
 Resistance
 Resistance
=
=
=
> 30M
> 30M
> 30M
Capacitance
Capacitance
Capacitance
=
=
=
57.86 nF
< 200 pF
< 200 pF
(T - R) Equivalent Cable Length =  3916.4 ft     based on 24 AWG PIC cable.
(T - G) Equivalent Cable Length =  N/A       based on 24 AWG PIC cable.
(R - G) Equivalent Cable Length =  N/A      based on 24 AWG PIC cable.
FIGURE 8.7.2A - RESULTS OF A AUTO-12 DMM TEST 
NOTE: Warning messages may appear at the bottom of the DMM result screen 
signifying that there is additional T-R capacitance, a missing shield or ground 
connection or grounded Tip or Ring wires. 
NOTE: For manually executed resistances tests, the maximum resistance will be up to 
999 Mohm 
For all tests, a user will have to press the RUN / STOP button to halt operation since all tests monitor the link 
continuously for changes except for the Autotests and the Capacitance test. 
Multiple tests results can be seen on the screen provided that NEXT TEST or DMM TESTS buttons are  
selected to change the test type to another test rather than pressing the function button for MAIN MENU. If a 
user presses MAIN MENU before a result(s) is saved or uploaded to a PC, all information on the screen will 
be lost. NEXT TEST and DMM TESTS remember the last test result so another test can be run without 
losing any information. 
At the completion of a test, the function button command selection at the bottom of the DMM TESTS 
RESULT screen consists of several options available to the operator. The following are brief descriptions of 
each of the function button options: 