User Manual
CableSHARK P3 User Guide 
117
12:58
NOV21
MAIN
MENU
E
CHG
100
10
1
Hits
F
SAVE/
UPLOAD
100
10
1
Hits
13:08
NOV21
13:18
NOV21
13:28
NOV21
Impedance (Terminated)
Noise Filter Selected
Impulse Threshold Level
Plot Interval
Impulse Test Duration
=
=
=
=
=
100 ohms
No filter
-43 dBm
1 minute
1 hour
Start Time
Cursor Time
Impulse Noise Count
Nov 21, 13:52
Nov 21, 14:04
15
*
Press  or  to move cursor
FIGURE 8.4.2E – IMPULSE HISTOGRAM RESULTS 
At the completion of a test (press FREEZE button to stop PSD tests), the function button command 
selection at the bottom of the Noise Test Result screen consists of several options available to the operator.  
The following are brief descriptions of each of the function button options: 
SHOW REF 
Pressing the function button corresponding to SHOW REF allows a user to 
overlay a REFERENCE MASK over the test result on screen to compare the 
two. See Section 5.4 for more information about the REFERENCE MASK 
feature. 
Pressing the SAVE / UPLOAD button shows a pop-up window where the user is allowed to upload results, 
and save results to internal memory or to external USB memory. 
UPLOAD RESULT 
Choosing UPLOAD RESULT from the SAVE / UPLOAD pop-up menu sends 
the result that appears on the CableSHARK screen (current test or recalled 
test) to a PC via a NULL modem cable. The remote control program Visi-
SHARK can be used to capture and analyze the results on a PC. 