tat2000instrev6.qxd 2/6/2006 EXERGEN 9:35 AM Page 1 Temporal Artery Thermometer Instructions For Use TemporalScannerTM 2000C Quick Check-List: - Read instructions completely before using - See www.temporalscanner.com for Educational Video - Remove protective cap before using - Reads arterial temperature, which is a core temperature very close to rectal temperature (See pg. 9) - Sensor should be clean (See pg. 12) - If perspiration is present see pg.
tat2000instrev6.qxd 2/6/2006 9:35 AM Page 2 Important Safety Instructions READ ALL INSTRUCTIONS BEFORE USING When using the product, especially when children are present, basic safety precautions should always be followed, including the following: z z z z z z z z z z z z z z z This product is intended for household use only. For information on thermometers for professional use, please see www.exergen.com, or call 617-923-9900. Use this product only for its intended use as described in this manual.
tat2000instrev6.qxd z z 2/6/2006 9:35 AM Page 3 If your thermometer will not be used regularly, remove the battery to prevent possible damage due to chemical leakage. If the battery leaks, remove carefully. Do not allow bare skin to touch leaking fluid. Dispose of used batteries properly. Do not wrap them in metal or aluminum foil. Wrap them in newspaper before disposing of them. Do not burn them. Battery may explode if overheated. SAVE THESE INSTRUCTIONS.
tat2000instrev6.qxd 2/6/2006 9:35 AM Page 4 The TemporalScanner has patented software, providing arterial heat balance. This unique process determines temperature by accurately measuring the balance between the tissues warming from arterial blood and tissues cooling/ warming caused by heat loss/gain to the environment.
tat2000instrev6.qxd 2/6/2006 9:35 AM Page 5 Before Using, Familiarize Yourself with the Instrument Protective Cap (remove during use) Probe Cone Probe Sensor Lens Probe LCD Display Screen LED Indicator Light SCAN Button Automatic turn-off in 30 seconds LED Indicator Light Thumb-Push Ridges Battery Compartment 9-volt alkaline battery z To Scan: Depress the button. The instrument will continually scan for the highest temperature (peak) as long as the button is depressed.
tat2000instrev6.qxd 2/6/2006 9:35 AM Page 6 Measuring TA Temperature What you should know before using the TAT: z Measure only the side of the head exposed to the environment. Anything covering the area to be measured (hair, hat, wig, bandages) would insulate the area, resulting in falsely high readings. z Slide the thermometer straight across the forehead, not down the side of the face.
tat2000instrev6.qxd 2/6/2006 9:35 AM Page 7 To ensure that the reading always reflects the body temperature accurately, you need to take account of the following factors which may affect an accurate reading. Sweating: When a fever resolves, your body may bring its temperature down by sweating. The TemporalScanner detects this reduction in temperature immediately - long before a rectal thermometer can do so. However, sweating also causes extra cooling of the skin.
tat2000instrev6.qxd 2/6/2006 9:35 AM Page 8 How to Take a Temperature 1. Remove protective cap before use. Be sure lens is clean. If not, clean with cotton swab dipped in alcohol and let dry. Hold the thermometer as shown. 2. Gently position the probe flush (flat) on the center of the forehead, midway between the eyebrow and the hairline. Press and hold the SCAN button. 3. Lightly slide the thermometer across the forehead keeping the sensor flat and in contact with the skin until you reach the hairline.
tat2000instrev6.qxd 2/6/2006 9:35 AM Page 9 Clinical Temperature Information Normal Body Temperature (BT) Normal BT is not a single temperature, but a range of temperatures influenced by age, time of day, and measurement site. You can establish your family’s normal ranges by taking a number of temperatures from each member during a day and keeping records of them. Many people may not have an elevated temperature even if they are ill.
tat2000instrev6.qxd 2/6/2006 9:35 AM Page 10 On a stable resting individual, temporal artery temperature is about the same as a rectal temperature, and approximately 0.8°F (0.4°C) higher than an optimum oral temperature. However, during fever episodes, the difference can be much higher, mainly because of the much greater speed of the TA compared to ear, oral, or rectal sites in responding to change in fever.
tat2000instrev6.qxd 2/6/2006 9:35 AM Page 11 Rectal Temperature Artifacts: Rectal temperature should only be considered as a good approximation of core temperature when the patient’s thermal balance is stable,8 and may be misleading after antipyretics, physical exercise, or other intervention that may change temperature quickly.
tat2000instrev6.qxd 2/6/2006 9:35 AM Page 12 Cleaning the Instrument The TemporalScanner is an optical instrument. Like a camera or eye glasses, a dirty lens will distort the view. If the thermometer is unable to see the heat clearly, it will be unable to measure it accurately, resulting in low readings. z Probe lens and cone should be shiny clean, if not, wipe with a small cloth or swab moistened with alcohol. z Hold upside-down to prevent excess moisture from entering the sensor area.
tat2000instrev6.qxd 2/6/2006 9:35 AM Page 13 Changing the Battery Blinking battery icon with temperature displayed: battery is low but will still operate correctly. Replace soon. Blinking battery icon with 2 dashes: not enough energy in the battery to measure correct temperature. Replace battery. Remove the battery compartment door by pushing down on the ridges with your thumb, and pushing away as indicated. Use both thumbs, if necessary.
tat2000instrev6.qxd 2/6/2006 9:35 AM Page 14 Display Messages A flickering Scn on display is visible during measurement. At completion, releasing the button will display and lock temperature on the screen for 30 seconds. The target temperature measured is higher than 107.6°F (42°C). The target temperature measured is lower than 60°F (15.5°C). Temperature of the thermometer is higher than 104°F (40°C). Let the instrument acclimatize for about 30 minutes in a cooler area in which it will be used.
tat2000instrev6.qxd 2/6/2006 9:35 AM Page 15 Product Specifications Clinical Accuracy Meets ASTM E1965-98 and EN60601-1 standards for electronic and radiation thermometers to the extent applicable to thermometers which measure the surface of the skin over the temporal artery. Regulatory Approvals CE Mark to -0197, TUV, Declaration of Conformity-ISO 9003/08.94, NIST certifiable traceable calibrations, UL listed.
tat2000instrev6.qxd 2/6/2006 9:35 AM Page 16 One Year Warranty Exergen Corporation warrants each new Exergen TemporalScanner 2000C (except battery) against defects in materials or workmanship for a period of one year from the date of purchase, and agrees to repair or replace any defective product without charge.