Specifications
EXCELL PRECISION CO., LTD.
SK-VET V1.0 SME300000067
5
Selectable Weight Units and Piece Unit:
Kilogram (kg) 1 g = 0.001 kg
Gram (g) 1 g = 1 g
Pound (lb) 1 g = 0.002204623 lb
Pound/Ounce
(lb, oz) 1 g = 0.03527396 oz
Ounce (oz) 1 g = 0.03527396 oz
Error Messages:
⇒ The EEPROM memory is not working correctly.
⇒ Zero is higher than the zero range when switching the scale on.
⇒ Zero is lower than the zero range when switching the scale on.
⇒ A/D value is unstable. When switching the scale on or pressing the ON/ZERO
key
or the TARE key, the A/D value is unstable for at least 20 seconds.
⇒ A/D value is less than -31250.
⇒ A/D IC malfunction (cannot read A/D value). The load cell may not be connected to
the scale correctly.
⇒ A/D value is higher than +31250.
⇒ The weight of the object is over the maximum capacity + 9 divisions.
⇒ After calibration the displayed resolution is greater than the internal resolution.
⇒ The gravity value is outside the range 9.772 ~ 9.829.
⇒ Error setting.
⇒ The weight value is greater than the zero range, when re-zeroing the
scale.
⇒ The weight value is below the zero range, when re-zeroing the scale.