Specifications
EXCELL PRECISION CO., LTD.
SK-VET V1.0 SME300000067
19
Use the
ON/ZERO
key
to
enter a parameter
Use the
ON/ZERO
key
to
enter a parameter
CHAPTER 5 ADVANCED FUNCTIONS
5-1. Auto Power off and Backlight Settings
Switch on the scale, press the
ON/ZERO
and the
HOLD
keys together to enter the advanced
function settings mode. The display shows 。
ON/ZERO ⇒ Increment digit (from 0 to 9)
TARE ⇒ Confirm / enter key
HOLD ⇒ Move the cursor to the right
Automatic Power-off Function
When the weight on weigh pan is less than 10d or remains steady for the set time, the scale
will automatically switch off.
v Automatic power-off is not functional when using an ac adaptor as the power source.
Therefore, auto power-off is active ONLY when using dry batteries as the power source.
Auto Backlight Function
When the weight is over 10d, the backlight is on. After the weight is stable for 10 seconds or
when the scale returns to zero, the backlight switches off.
The display shows the current setting
Auto power off timer setting
Use the
ON/ZERO
key to enter a parameter.
0 ⇒ No auto power-off
1 ⇒ When the scale is idle for 1 minute, the
scale automatically switches off
2 ⇒ When the scale is idle for 2 minutes, the
scale automatically switches off
19 ⇒ When the scale is idle for 19 minutes,
the scale automatically switches off
Press the
TARE
key
The display shows the current setting
Auto backlight setting
Use the
ON/ZERO
key.
on ⇒ Auto Backlight on
off ⇒ Auto Back light off
NEXT