Specifications
EXCELL PRECISION CO., LTD.
SK-VET V1.0 SME300000067
18
Key in the local gravity
value
4-5.Gravity Compensation
Set the jumper SWA1 on the main board to the ADJ position. Press the ON
/ZERO
key to switch on the scale, and press the ON
/ZERO
key again before the display
resets back to zero. The scale enters the gravity compensation mode.
Set the jumper SWA1 to the LOCK position
The scale returns to the weighing mode
v The gravity value must lie between those at the Equator and the Polar Regions.
Gravity value at the equator G
E
= 9.7803184558 m/sec
2
Gravity value at the polar region G
P
= 9.8321772792 m/sec
2
Taipei: 9.78914 m/sec
2
Shanghai: 9.79585 m/sec
2
v The display shows error message
, when the gravity value is NOT within the
range (9.829 > gravity >9.772).
.
.
ON/ZERO ⇒ Increment digit (from 0 to 9)
TARE ⇒ Confirm / enter key
HOLD ⇒ Move the cursor to the right
.
Press the
TARE
key
The display shows and saves
the entered gravity value in
the EEPROM.
The display shows the current