Place a load on the weigh
pan according to the weight
value show on the display,
then press
the
TARE
key
.
Remove the load from the
weigh pan.
4-4.Span Calibration
v Set the jumper SWA1 on the main board to the ADJ position (EEPROM
UNLOCKED). Press the ON
/ZERO
key to enter the SPAN calibration mode;
the display shows .
v Set the jumper SWA1 back to the LOCK position, after completing the
calibration.
v If the jumper SWA1 is set back to the LOCK position during calibration, the
scale exits the service mode automatically.
Set the jumper SWA1 to the LOCK position.
The scale returns to the weighing mode.
.
.
.
ON/ZERO Increment digit (from 0 to 9)
TARE Confirm / enter key
HOLD Move the cursor to the right
.
Press the
TARE
key
Detect and save zero point in EEPROM
automatically.
The maximum capacity is displayed.
The calibration data is saved in the
EEPROM