CHAPTER 4 SERVICE MODE ACCESS
v Set the jumper SWA1 on the main board to the ADJ position (EEPROM
UNLOCKED). Press and hold the TARE key, followed by pressing the
ON
/ZERO
key. Release the ON
/ZERO
key first and then the TARE key to
enter the service mode; the display shows .
v Set the SWA1 jumper back to the LOCK position when service configuration is
completed.
v If the SWA1 jumper is set to its LOCK position during calibration, the scale exits
the service mode automatically.
Capacity Configuration
Weight Calibration
Function Settings
Press the
ON/ZERO
key
Press the
UNITS
key
Press the
ON/ZERO
key
Press the
ON/ZERO
key
Press the
UNITS
key
Press the
UNITS
key