EXCELL PRECISION CO., LTD. © Excell Precision Limited 2005. All rights reserved Worldwide. The information contained herein is the property of Excell Precision Limited and is supplied without liability for errors or omissions. No part may be reproduced or used except as authorised by contract or other written permission. The copyright and the foregoing restriction on reproduction and use extend to all media in which the information may be embodied. SK-VET V1.
EXCELL PRECISION CO., LTD. TABLE OF CONTENTS BEFORE USING THE SCALE.......................................................................................... 2 SAFTY INSTRUCTIONS................................................................................................... 2 PREPARING TO USE THE SCALE.................................................................................. 3 CHAPTER 1 FEATURES AND SPECIFICATIONS .......................................................... 4 CHAPTER 2 DISPLAY ...
EXCELL PRECISION CO., LTD. BEFORE USING THE SCALE Thank you for purchasing an EXCELL Electronic Digital Scale. In order to use the scale properly, please read this User Manual carefully before use. If you have a problem concerning the scale, please contact your supplier. 1. Please keep scale in a cool and dry place. Do not store at high temperatures. 2. Do not allow any liquids to come into or contact with the scale. If it is happened, wipe the scale dry with a cloth. 3.
EXCELL PRECISION CO., LTD. PREPARING TO USE THE SCALE 1. For accurate weight readings locate the scale on a firm level surface free from vibrations. 2. Avoid operating the scale in direct sunlight or drafts of any kind. 3. Remove any weight that might be on the weigh pan before the scale is switched on and avoid leaving weight on the pan for long periods of time. 4. Once the scale has been switched on, it will go through a LCD display test and then re-zero to be ready for use. 5.
EXCELL PRECISION CO., LTD. CHAPTER 1 FEATURES AND SPECIFICATIONS Features: n Multi-function operation: Fast weighing operation; Full range tare; Auto power-off; Weight hold feature. n Dual-weighing units: Pound (lb) and kilogram (kg). n User-friendly design: Auto calibration; Large LCD display (digit height 12mm x 25mm); Gravity compensation; Low battery warning indicator; Double over-load protection. n Dual-power source: The power source can either be from AC / DC or dry batteries.
EXCELL PRECISION CO., LTD. Selectable Weight Units and Piece Unit: Kilogram (kg) Gram (g) Pound (lb) Pound/Ounce (lb, oz) Ounce (oz) 1g 1g 1g 1g 1g = = = = = 0.001 kg 1g 0.002204623 lb 0.03527396 oz 0.03527396 oz Error Messages: ⇒ The EEPROM memory is not working correctly. ⇒ Zero is higher than the zero range when switching the scale on. ⇒ Zero is lower than the zero range when switching the scale on. ⇒ A/D value is unstable.
EXCELL PRECISION CO., LTD. CHAPTER 2 DISPLAY 1 2 3 TARE: “Tare” indication. ZERO: “Zero” indication. ► 1: Kilogram (kg), the 1st unit indication. ► 2: Pound (lb), the 2nd unit indication. ► 3: Hold mode indication. : “Low Battery” indication. SK-VET V1.
EXCELL PRECISION CO., LTD. CHAPTER 3 KEYBOARD FUNCTION 1. 2. OFF : The key to switch the scale off. ON ZERO : ON/ZERO key. When the scale is off, press the key to switch it on. When the scale is switched on, with the weigh pan empty, if the display is not showing zero, press this key to zero (balance) the scale. v Zero range: ± 2% of full scale. TARE 3. : TARE key. Press the key to deduct the weight of the container.
EXCELL PRECISION CO., LTD. CHAPTER 4 SERVICE MODE ACCESS v Set the jumper SWA1 on the main board to the ADJ position (EEPROM UNLOCKED). Press and hold the TARE key, followed by pressing the key. ON/ZERO Release the ON/ZERO key first and then the TARE key to enter the service mode; the display shows . v Set the SWA1 jumper back to the LOCK position when service configuration is completed.
EXCELL PRECISION CO., LTD. 4-1. Capacity Configuration ON/ZERO Press the TARE key ⇒ Increment digit (from 0 to 9) TARE ⇒ Confirm / enter key HOLD ⇒ Move the cursor to the right Input the parameter “01” to enter the capacity configuration The display shows the maximum capacity + 1 division and the weight unit set last time for 2 seconds. Press the TARE key . ► Weight unit or piece unit setting Example: Two weight units “kg” & “lb”, with “kg” as the units used for weight calibration.
EXCELL PRECISION CO., LTD. Refer to “Note 3” Example: 1 ⇒ Division size 1 ⇒ The number of decimal places 0 ⇒ range setting Press the TARE key The display shows the max. capacity of the scale + minimum division and the unit weight for 2 seconds. . Input the parameter “00” Press the TARE key to complete the capacity configuration SK-VET V1.
EXCELL PRECISION CO., LTD. NOTE 1 The users can set up the different weight units in various orders according to their preference, and the amount of the chosen weight units can be up to 4. (a) (b) (c) (e) (a) ⇒ The first weight unit (only “kg”, or “lb” are available to choose from). (b) ⇒ The second weight unit (select one of the parameters listed below). (c) ⇒ The third weight unit (select one of the parameters listed below). (e) ⇒ The amount of the weight units selected (select one of parameters 1 ~ 3).
EXCELL PRECISION CO., LTD. 4-2. Weight Calibration ON/ZERO Press the TARE key ⇒ Increment digit (from 0 to 9) TARE ⇒ Confirm / enter key HOLD ⇒ Move the cursor to the right Input the parameter “01” The display shows the calibration weight value and the weight unit set at the first position. Press the TARE key ► . Unload the weigh pan Press the TARE key When the weight reading is stable the display shows the calibration weight value.
EXCELL PRECISION CO., LTD. 4-3. Function Settings Press the TARE key Input a parameter 00, 01, or 02. After inputting “00”, Press the TARE key ON/ZERO to return to the upper level TARE ⇒ Confirm / enter key HOLD ⇒ Move the cursor to the right ⇒ Increment digit (from 0 to 9) ⇒ Return to the upper menu level ⇒ Environment settings ⇒ Buzzer type SK-VET V1.
EXCELL PRECISION CO., LTD.
EXCELL PRECISION CO., LTD. Stable range parameter setting The display shows the current setting. Cont.
EXCELL PRECISION CO., LTD. 4-3-2 Buzzer Type Press the TARE key Buzzer type The display shows the current setting Use the ON/ZERO key to Buzzer type Use the ON/ZERO key to key in a parameter 0 ⇒ Pulse signal (AC buzzer, PT-134OP, 4KHz) key in a parameter 1 ⇒ Continuous signal (DC buzzer, UIC-4201) Press the TARE key Key in the parameter “00” ON/ZERO Press the TARE key SK-VET V1.
EXCELL PRECISION CO., LTD. 4-4.Span Calibration v Set the jumper SWA1 on the main board to the ADJ position (EEPROM UNLOCKED). Press the ON/ZERO key to enter the SPAN calibration mode; the display shows . v Set the jumper SWA1 back to the LOCK position, after completing the calibration. v If the jumper SWA1 is set back to the LOCK position during calibration, the scale exits the service mode automatically. Press the TARE key Detect and save zero point in EEPROM automatically.
EXCELL PRECISION CO., LTD. 4-5.Gravity Compensation Set the jumper SWA1 on the main board to the ADJ position. Press the ON/ZERO key to switch on the scale, and press the ON/ZERO key again before the display resets back to zero. The scale enters the gravity compensation mode. The display shows the current compensation value . Key in the local gravity value . Press the TARE key The display shows and saves the entered gravity value in the EEPROM. ON/ZERO .
EXCELL PRECISION CO., LTD. CHAPTER 5 ADVANCED FUNCTIONS 5-1. Auto Power off and Backlight Settings Switch on the scale, press the ON/ZERO and the HOLD keys together to enter the advanced function settings mode. The display shows 。 The display shows the current setting Auto power off timer setting Use the ON/ZERO key to enter a parameter.
EXCELL PRECISION CO., LTD. Cont.
EXCELL PRECISION CO., LTD. NOTE 4 parameter setting (Animal Weighing mode) (a) (b) (c) (d) (e) (a) ⇒ Parameters 0~4 to set the stabilization timer 0: 0.5 second 1: 1 second 2: 2 seconds (default setting) 3: 4 seconds 4: 8 seconds (b) (c) (d) (e) ⇒ Dynamic stabilization range setting: 0 ~ 100% (pre-set as 5%) Example: To set the stabilization timer as 4 seconds and the range as 1.5%, key in 30015. SK-VET V1.
EXCELL PRECISION CO., LTD. APPENDIX 7 SEGMENT DISPLAY CHARACTERS: 7 segments Alphabet Digit letter 7 segments letter Alphabet 0 A N 1 B O 2 C P 3 D Q 4 E R 5 F S 6 G T 7 H U 8 I V 9 J W K X L Y M Z SK-VET V1.