 A ⇒ 0 = Buzzer beeps when being stable; LCD indication lights up. 
 1 = Buzzer beeps when being stable; LCD indication lights up without stable 
determination. 
 2 = Buzzer beeps and LCD indication lights up without stable determination. 
 B ⇒ 0 = No buzzer beep. 
 1 = When it’s OK (the value higher than lower limit and lower than upper limit,) 
the buzzer beeps. 
 2 = When the value is lower than or equal with lower limit, or higher than upper 
limit, the buzzer beeps. 
4 Single Point Presetting (lower limit only), the buzzer is fixed as: 
  When the value is higher than lower limit, the buzzer beeps and 
LOW indication will be off. 
 Buzzer beeps and LCD indication lights up without stable determination. 
4 Single Point Presetting (upper limit only), the buzzer is fixed as: 
  When the value is higher than upper limit, the buzzer beeps 
and HIGH indication will be on. 