- EXCELL PRECISION Weighing Scale User Manual
 EXCELL PRECISION CO., LTD. 
Dolphin 9903 V 1.1                                          SME300000041 
18
Put the platter on and make sure 
there is no weight on the platter. 
Then press the  TARE  key. 
Place enough weight to the platter 
according to the display. 
When the scale becomes stable, the 
buzzer beeps for 3 times. 
Remove all the weights, and the 
calibration procedure is completed. 
     .     
     .     
2. External Weight Calibration   
4         is not accessible for approval models. 
4  It is recommended that the external weight calibration should be carried out by a 
qualified technician from your supplier to ensure the accuracy of the scale. 
 ON/ZERO  ⇒ Upward digit selecting (0-9) 
 UNITS   ⇒ Downward digit selecting (9-0) 
 TARE   ⇒ Move the cursor rightward 
 NET/GROSS   ⇒ Move the cursor leftward 
Press the 
 TARE 
 key
The display shows the calibration 
weight (
equals to the full capacity) and 
the weight unit. The last digit flashes. 
4 The calibration weight can be 
modified if necessary. 
kg 
kg 
kg 
The scale is calculating the internal 
value of the zero point. 
The scale becomes stable
The display shows the calibration 
weight
. 