User's Guide

User Guide
- 40
26. Set exposure time
Command to S2000 PRO/ELITE; STMn
Note: 2 ≤ n ≤ 9999
n must be a integer number
Example:
If “STM10XX\r” is set, the S2000 PRO/ELITE exposure
time will be 1 second
Response from S2000 PRO/ELITE;
Received if the value n is valid
Invalid if the value n is out of range
EXP the S2000 PRO/ELITE is running a timed exposure,
so the value cannot be set.
27. Set LEVEL set point
Command to S2000 PRO/ELITE; SPWn
Note: n ≥ .1 (5mm LG)
n ≥ .25 (8mm LG) and n is a float number
Example:
If “SPW2.55XX\r” is set, the S2000 PRO/ELITE LEVEL
set point will be 2.55 W.
Response from S2000 PRO/ELITE;
Received if the value n is valid
Invalid if the value n is out of range
Need CAL the S2000 PRO/ELITE need to be calibrated
before the value can be set.
EXP- the S2000 PRO/ELITE is running a timed exposure,
so the value cannot be set.
28. Set irradiance set point
Command to S2000 PRO/ELITE; SIRn
Note: n > 0, and n is a float number
Example:
If “SIR15.32XX\r” is set, the S2000 PRO/ELITE
irradiance set point will be 15.32 W/cm2.
Response from S2000 PRO/ELITE;
Received if the value n is valid
Invalid if the value n is out of range