User's Guide

User Guide
- 39
Example:
If “15.66XX\r” is responded, the S2000 PRO/ELITE
irradiance set point will be 15.66 W/cm2.
If the unit is in LEVEL mode, “Need CAL” will be the
response.
23. Get S2000 PRO/ELITE actual irradiance reading
Command to S2000 PRO/ELITE; GIM
Response from S2000 PRO/ELITE; n (n represents a float number
in Watts per square cm)
If the unit is in LEVEL mode, “Need CAL” will be the response.
Example:
If “15.66XX\r” is responded, the S2000 PRO/ELITE
irradiance reading will be 15.66 W/cm2.
24. Get S2000 PRO/ELITE maximum irradiance output
Command to S2000 PRO/ELITE; GMP
Response from S2000 PRO/ELITE; n (n represents a float number
in Watts per square cm)
If the unit is in LEVEL mode, “Need CAL” will be the response.
Example:
If “35.66XX\r” is responded, the S2000 PRO/ELITE
maximum irradiance output will be 35.66 W/cm2.
25. Set iris Level
Command to S2000 Pro/Elite; SILn
Note: 0 < n ≤100
n must be a integer number
Example:
If “SIL20XX\r” is set, the S2000 PRO/ELITE iris Level will
be 20%.
Response from S2000 PRO/ELITE;
Received if the value n is valid
Invalid if the value n is out of range
Done CAL the S2000 PRO/ELITE has been calibrated,
so the value can not be set.
EXP the S2000 PRO/ELITE is running a timed exposure,
so the value cannot be set.