If “10XX\r” is responded, the software version on the I/O
board will be 1.0.
19. Get S2000 PRO/ELITE serial number
Command to S2000 PRO/ELITE; GSN
Response from S2000 PRO/ELITE; n (n represents a integer
number)
Example:
If “20XX\r” is responded, the S2000 PRO/ELITE serial
number will be 20.
20. Get light guide diameter
Command to S2000 PRO/ELITE; GLG
Response from S2000 PRO/ELITE; n (n represents a float number)
Example:
If “2.5XX\r” is responded, the light guide diameter will be
2.5mm
Note:
If “3.402823466e38XX\r” is responded, no light guide
diameter information is available.
If the unit is in LEVEL mode, “Need CAL” will be the
response.
If “-3.403823466e38XX\r” is responded, no light guide
diameter information is available. The S2000 PRO/ELITE
displays LEVEL instead of irradiance.
21. Get S2000 PRO/ELITE POWER set point
Command to S2000 PRO/ELITE; GPW
Response from S2000 PRO/ELITE; n (n represents a float number
in Watts)
Example:
If “1.222XX\r” is responded, the S2000 PRO/ELITE
LEVEL set point will be 1.222 W.
If the unit is in LEVEL mode, “Need CAL” will be the
response.
22. Get S2000 PRO/ELITE irradiance set point
Command to S2000 PRO/ELITE; GIR
Response from S2000 PRO/ELITE; n (n represents a float number
in Watts per square cm)