User's Guide

User Guide
- 36
If “50XX\r” is received, the unit status is as follows:
Alarm - Off; Shutter Home Pass; Lamp On; Lamp
Ready Yes; Shutter Close; Front Panel locked;
Calibration No; Exposure Fault No.
7. Run a timed exposure
Command to S2000 PRO/ELITE; RUN
Response from S2000 PRO/ELITE; Received
8. Open Shutter
Command to S2000 PRO/ELITE; OPN
Response from S2000 PRO/ELITE; Received
9. Close Shutter
Command to S2000 PRO/ELITE; CLS
Response from S2000 PRO/ELITE; Received
10. Turn On Lamp
Command to S2000 PRO/ELITE; TON
Response from S2000 PRO/ELITE; Received
11. Turn Off Lamp
Command to S2000 PRO/ELITE; TOF
Response from S2000 PRO/ELITE; Received
12. Get Lamp Configuration
Command to S2000 PRO/ELITE; GLH
Response from S2000 PRO/ELITE; n (n represents a integer
number)
Bit 15: Abuse Bit (0=OK, 1=lamp abused)
Bit 14: Lamp Type (0= surface curing, 1=standard)
Bit 13 Bit 0: Lamp Hours
Example:
If “20XX\r” is responded, the lamp configuration is as
follows:
No abuse bit set, surface curing with 20 hours lamp.
13. Get calibrated lamp hours
Command to S2000 PRO/ELITE; CLH
Response from S2000 PRO/ELITE; n (n represents a integer
number in hours)
Note: Please refer to Table 6 Unit Status for details.