5.3 Safety Precautions
The S2000 PRO/ELITE is equipped with two safety sensors to protect the user
from accidental UV exposure. In addition, please observe the following
precautions during use. This Series of cautions, warnings and dangers relate to
the operation and maintenance of the S2000 PRO/ELITE. They are also
presented throughout this User’s Guide where necessary.
Warning
Eye damage may result from directly viewing the light produced by the lamp used
in this product. Always use the UV protective eyewear supplied with the unit and
always turn the lamp off before removing lamp housing cover.
Warning: UV protective eyewear must meet the following recommended optical
specifications:
Spectral range; 320-500 nm
WARNING: Do not stare directly at the light emitted from the lamp. This may be
harmful, resulting in eye injury. Always use UV protective eyewear as indicated
below. Additionally, protect any exposed skin with appropriate clothing or shielding
as required.
Caution
Never look into the light emitting end of the light guide. The light could severely
damage the cornea and retina of the eye if the light is observed directly. Eye
shielding must be used at all times as well as clothing to protect exposed skin.
Warning
Always make sure the light guide is properly inserted into the S2000 PRO/ELITE
prior to turning on POWER to the unit. This will minimize the risk of exposure to the
UV light.
Warning
To reduce the risk of fire or shock, always replace the fuses with the same type
and rating. Always detach the POWER supply cord prior to attempting to replace
fuses!
Warning
Disconnecting of the main supply source is only possible by unplugging the
POWER cord.