Operator Manual Corival V3 English Lode BV Groningen, The Netherlands 0344
Operator manual Corival This is the short operator manual for the Lode Corival Ergometer, which describes the basic actions for starting up your ergometer. For a complete description, we refer to our User Guide. Intended Use A Lode ergometer is a diagnostic tool intended to be used as a stress test device in a medical environment. The main goal of the use of a Lode ergometer is to create reproducible stress tests.
Operator manual Corival Caution: Not suitable for use in the presence of flammable anaesthetics. Note: The voltage selector (115/230VAC) is inside the ergometer. The main unit panel has to be unscrewed. Voltage selection should be done by a technician! Possible hazards Using the Corival according to intended use, contra indications, maintenance, precaution and common sense stated above may not eliminate all hazards.
Operator manual Corival TABLE OF CONTENTS 1 DESCRIPTION OF THE CORIVAL ERGOMETER 1.1 1.2 1.3 MANUAL CONTROL QUICK START ANALOG EXTERNAL CONTROL TERMINAL – DIGITAL EXTERNAL CONTROL VIA RS232 2 MAIN MENU 3 SYSTEM PARAMETER 3.1 3.1.1 3.1.2 3.1.3 3.1.4 3.1.5 3.1.6 3.2 3.3 3.3.1 3.3.2 3.3.3 3.3.4 3.3.5 3.4 3.4.1 3.4.2 3.4.3 3.4.4 3.4.5 3.4.6 3.4.7 3.4.8 4 SPECIFICATIONS 4.1 4.2 5 TECHNICAL SPECIFICATIONS FUSES COMMUNICATION PORTS 5.1 5.2 5.3 5.
Operator manual C.2 C.3 D Using the Heart rate option Installing the BP cuff Blood pressure option 928817, -818, 820 D.1 D.2 D.2.1 D.2.2 D.3 D.4 D.5 D.6 E 28 28 29 29 29 31 31 32 33 34 0 Watt Start-up System 906805 36 F.1 F.2 G Using the BP system Patient Hook-up Apply the cuff Position of the cuff Abort codes BP system Cleaning and disinfection Cleaning of the cuff: Installing the BP cuff 27 27 SpO2 option 928840 E.1 E.
Operator manual Corival 1 Description of the Corival ergometer The control panel consists of the following elements: 1. 2. 3. 4. 5. 6. Led indicator for rpm. Display A directed to test subject. Down button. Up button. Confirmation button. Ergometer type indication. Fig. 1 Control Panel standard Corival The back panel of the Corival contains the following elements: 1. 2. 3. 4. 5. 6. 7. 8. 9.
Operator manual Corival Adjustability The ergometer is adjustable to test subjects of most sizes and ages. • The height of the saddle can be adjusted by tilting the handle on the back underneath the saddle • The height of the handlebars can be adjusted by loosening the handle in front of the ergometer. Moving the ergometer The ergometer can be moved by turning it over to the right (appr. 30º) when you are in front of the handlebar. A wheel in the middle of the ergometer drops out.
Operator manual 1.2 Corival Analog external control Execute the following steps to control the Corival externally using the analog in-/output: 1. Connect the external controling device (e.g. ECG system) with a correct cable to the analog i/o port. 2. Start the ergometer and go in the main menu to ANALOG. 3. Now the ergometer can be controlled by an external device with an analog signal. 1.
Operator manual Corival 2 Main menu When the ergometer is started after pressing the confirmation button ( ) you enter the MAIN MENU. The following submenus are selectable from the main menu: ! TERMINAL ! MANUAL ! ANALOG ! SYSTEM PARAMETER Use the arrow keys▼▲ to scroll through the menus. Press the ( ) button to select a menu. Keeping the ( ) pressed for a few seconds allows you to go up one or more menu levels, wherever you are in the menu. The complete menu structure is shown below.
Operator manual Corival 3 System parameter In the System Parameters different specific ergometer settings can be carried out. 3.1 Set display This function allows you to adjust the parameters which can be read out on the display during exercising. If more then 2 are set to ON they are shown in groups of 2 scrolling. Choosing default allows you to go to factory settings for the display parameters.
Operator manual 3.3 Corival Set mode Set mode is used to define which parameter (like load, torque, heart rate and linear constant alpha) has to be adjusted during exercising in MANUAL mode with▼▲ keys 3.3.1 Hyperbolic Hyperbolic mode means that the load P can be adjusted in watt. The load is kept constant independent of rpm. 3.3.2 HR controlled Heart rate can be adjusted and the load is regulated until the adjusted heart rate is reached.
Operator manual 3.4.2 Corival Keyboard On/Off If you do not want to interrupt a training or you do not want the test subject to interrupt the training by pressing a key accidentally, the possibility is offered to disable the keyboard Now you wonder how to get back the control to the keyboard. Act as follows: Switch off the ergometer Switch it on again while keeping the ( ) button pressed. Now you have back the control. Go to SYSTEM PARAMETERS / KEYBOARD ON/OFF and set it to ON.
Operator manual Corival 4 Specifications 4.1 Technical Specifications Principle Mains voltage Frequency Patient capacity Power consumption Workload Device control Accuracy : : : : : : : : Read-out workload Read-out pedalling speed : : eddy current brake small disk. 115-230 VAC (adjustable) 50 Hz, 60 Hz. 350 lb - 160 kg. 120 VA. 0-999 W HYP, 0-999 W LIN manual, external analog, external digital < 3 W for P <= 100 W, < 3% for 100 W< P<500W, <5% for P>500 W. digital, 0-999 resolution 1 W.
Operator manual 5 Corival Communication ports This chapter describes the communication ports in the main unit panel of the Corival: ! RS232 (two ports for RS232-in and loop connection RS232-out) ! Analog in/out 5.1 RS232 The Corival has two RS232 connectors for loop-connecting a number of Corival’s and for communication with a personal computer or an ECG device. When the ergometer is controlled externally always use the RS232-in port.
Operator manual A Corival Maintenance The Lode Ergometer is mechanically and electronically designed for durability and accuracy. This also accounts for the options and add-ons of the device. The frequency of calibration depends on the intensity of usage. As a rule Lode Ergometers are calibrated on an annual basis. Please refer to the Service Manual for specific details on check-up intervals for the Lode Ergometer and its options and/or add-ons.
Operator manual B Corival Programmable Display Unit 928810 B.1 Installing the Programmable Display Unit (option 928810) When the option 928810 is ordered later on, it should replace the standard Control Unit. Replacement should be carried out by service personnel only. For the description, we refer to our user guide supplied with this ergometer. B.2 Control panel Corival with Programmable Display Unit (option 928810) 1. 2. 3. 4. 5. 6.
Operator manual Corival LODE P-display ver X.YZ In case the Blood pressure option is available it is followed for approximately 1 second by the text "BPM Initialized "thereby indicating correct operation of the BP device. The SpO2 comes together with the P-Display.
Operator manual Corival 3. TIMER 3. DISTANCE 3. TARGET HR* 3. ENERGY* 2. SET P-SLOPE 2. SET MODE 3. HYPERBOLIC 3. HR CONTROLLED 3. LINEAR 3. FIXED TORQUE 3. ISOKINETIC (only available for Excalibur Sport 925900) 2. SETTINGS 3. DEFAULT STARTMENU 3. ECG START* 3. P IN/OUT ANALOG 3. CONTRAST 3. ECG TRIGGER 3. TC_BYTES 3. RS232 PROTOCOL 3. LANGUAGE 3. DEVICE NO 3. SEATING POSITION (only available for Excalibur Sport 925900) 1. PROTOCOL* 2. PROTOCOL EDIT * 2. PROTOCOL RUN * 3. ÅSTRAND PROTOCOL* 2.
Operator manual Corival Notice: ♥ = 0/m* Only displayed if HR option is being installed! The ergometer is now ready for external control with an analog signal as previously described. See for more information about analog control the user guide (only available in English) supplied with the ergometer. B.3.
Operator manual Corival Where the function POWER is the same as described previously in this manual (power indication on or off), the button ESC will bring you back to the system parameters menu. With the ▼▲ keys you can select the power indication to be ON or OFF, and NEXT will bring you to the next item of the SET DISPLAY menu. Beside the parameters RPM, HEART RATE, TORQUE, TIMER and DISTANCE two more are parameters are added: TARGET HEART RATE and ENERGY.
Operator manual Corival accidentally, you get into to the TEST FUNCTIONS it can cause damage to the ergometer operation and guarantee will be lost! B.3.7 Protocol In this paragraph the program possibilities under the menu PROTOCOL will be described: protocol run PROTOCOL EDIT protocol delete prot edit You can select the menus PROTOCOL EDIT; PROTOCOL RUN and PROTOCOL DELETE. B.3.7.1 Protocol edit The menu PROTOCOL EDIT allows you to program new exercise protocols or edit an existing protocol.
Operator manual Corival changed, in the example mentioned above: S. Pressing the ▼▲ keys give you the ability to toggle between the possibilities for programming a stage in a protocol: S, T, L, H, P and Recovery. the workload can be adjusted + or – and the duration of this hyperbolic stage can be determined; the load is independent of the RPM. T= torque fixed torque: the torque in Nm can be adjusted with certain duration.
Operator manual Corival test Step P= 0W t=00:00 1S dP=+050W dt=060s esc next Confirm the S with ( ), use the ▼ key to decrease the workload to 10 watt. 7. Press the confirmation button NEXT until you reach the following screen. Adjust time dt to 60 seconds using the ▼▲ keys. test Torq P= 10W t=01:00 2T T= 10Nm dt=030s back next 8. 9. 10. 11. 12. Select T with the ▼▲ keys and adjust the Torque to 10 Nm and the duration to 30 seconds with the same keys. Press the confirmation button.
Operator manual Corival the total protocol and define the recovery stage before you leave this menu otherwise the changes will not be saved. B.3.7.3 Protocol Run The menu PROTOCOL RUN is used to select the desired protocol for the exercise test. Select Protocol Run and three previously programmed protocols will appear on screen. To scroll through the total list use the arrows, for instance: 10w min 25w 2 min ramp 400w esc ok If you want to select e.g.
Operator manual Corival 1 t=00:59 P= 10 W ♥ = 83/min rec θ=00:01 58 rpm rec In this example the screen shows that this is the first stage, t= 00:59 with 59 sec to go, the timer is showing 1 sec of total exercise time, the timer will stop the registration when the RPM drops below 30. The workload P is 10 watt, the pedaling frequency is 58 RPM, and the heart rate is 83 beats per minute. The three displayed parameters are those selected under SYSTEM PARAMETERS / SET DISPLAY.
Operator manual Corival Rec P= 50W P= 50W ♥ = 174/min esc t=10:30 60 rpm esc In this example the recovery phase will bring the workload down to 50 watt. The timer is showing 10:30 seconds and will go on with the registration during the recovery as long as the RPM is above 30 (the BPM can still be measured when BPM is displayed at the confirmation button). To stop the recovery phase press the ESC button to go back to the PROTOCOL menu.
Operator manual C Corival Heart rate option 928924 C.1 Connecting the Heart rate system (option 928924) To install a heart rate option purchased afterwards, follow the instructions given in the user guide supplied with the ergometer. Installation of the Heart rate option itself has to be done by service personnel only. C.2 Using the Heart rate option The heart rate option is automatically recognized by the ergometer.
Operator manual D Corival Blood pressure option 928817, -818, 820 With option 928817, -818, 820 it is possible to perform stand alone or external controlled blood pressure measurements. After installation the BP module is recognized automatically at start up of the ergometer, indicated by the following screen for a few seconds: LODE ERGOMETER TBD D.1 Using the BP system Caution: Carefully read the instructions in Appendix D.2 Patient Hook-up before continuing.
Operator manual Corival mm Hg below the last measured DIASTOLIC pressure. The used deflate rate is approximately 4 mm Hg/heartbeat. A list of most occurring warnings and errors is given in D.3 Abort codes BP system. The built-in BP module offers, in case of malfunction or poor performance, the possibility to make error codes visible. See for more information the user guide supplied with this ergometer. Caution: The interpretation of the error codes should be done only by experts.
Operator manual Corival Fig 3: Cuff position Caution: Never place the cuff so that the microphone is located in the elbow. Pull the cuff so tight, that it neither moves nor ties off the arm of the patient. By fasten the cuff too tight or too loose, error measurements are caused. Wrap the end of the cuff around the upper arm and close the Velcro fastening (Fig 4: Mounting the cuff). Fig 4: Mounting the cuff Ensure that the air hose from the monitor to the cuff is not compressed, crimped or damaged.
Operator manual Corival The intervals of the measurements should not be too short. Intervals of 2 minutes and more result in long pauses between the stenoses, so that venous congestion and artiospasm can be avoided. D.3 Prior. Abort codes BP system Error message Display Possible error cause 1 ACHECK ECG SIGNAL@@ 0 / 0 mmHg ! ! ! ! 2 ACUFF OVERPRESSURE@@ 0 / 0 mmHg ! Strong arm movements during the blood pressure measurement. ! Air hose is bent or blocked.
Operator manual Corival Entry- and connection cables, like the air-hose from the cuff should not be sterilized in hot steam or cold gas or plunged in disinfectants. D.5 Cleaning of the cuff: For the cleaning of the cuff, you can take out the air-hose (Fig 5: Air-hose removal). Fig 5: Air-hose removal Now you can open the Velcro fastening of the cuff and take out the microphone, Fig 6: Microphone removal. Fig 6: Microphone removal The cuff can be washed at 30° C.
Operator manual Corival Fig 7: Placing the air-hose Slip the microphone back in the bag (Fig 8: Placing the microphone) and close the Velcro fastening. Fig 8: Placing the microphone D.6 Installing the BP cuff To install a BP option purchased afterwards, follow the instructions given in the user guide supplied with the ergometer. Installation of the BP module itself has to be done by service personnel only.
Operator manual E Corival SpO2 option 928840 With option 928840 your Corival is able to perform stand alone or external controlled measurement of functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate (BPM) for adult and pediatric test subjects. Keep in mind the fact that SpO2 measurements during arm exercise is not possible due to movement artifacts. Do not attach the pulse oximeter to the same limb as the Cuff of the blood pressure monitor or any other blood flow restrictor.
Operator manual Corival TERMINAL P= 0W 0 rpm ♥= 55/min SpO2= SNSD ecs esc The following errors can be displayed: − SNSD Sensor Disconnect – An absence of signal, meaning the sensor is gone − OOT Out Of Track – An absence of consecutive good pulse signals − MIDA SpO2 could not be calculated, missing data indicator − ARTF Artifact – A detected pulse beat didn’t match the current pulse interval − MPRF Marginal Perfusion – Amplitude representation of medium signal quality (holds for entire duration) − LPRF Lo
Operator manual F Corival 0 Watt Start-up System 906805 With option 906805 your Corival is able to start the Corival without load (0 watt) at a preset pedal speed (rpm). You can choose any pedal speed between 30 and 80 rpm. The number of revolutions is increased from 0 to the required rpm by means of a motor drive. With this option there is no energy-loss due to the start-up phase. This option has to be built-in in the Corival while manufacturing and cannot be added later on. F.
Operator manual Corival 0 WATT STARTUP SYST. 50 rpm 0 rpm manual ↓ ↑ manua The default target rpm (50) is shown on the left. The current rpm is shown on the right. With the up and down key's the target rpm can be changed to the desired value. On the first button press the 0 Watt startup system starts, when the current rpm is above the target rpm the system is released and the display returns to the previous selection (Manual).
Operator manual G Corival RS232 Connections Your ergometer can be connected to almost all available external equipment. In the overview that follows, you can find an overview of the manufacturers of this equipment, the models and the protocol that should be selected in your ergometer. Please keep into account that this list is not exhaustive and that it is possible to connect your Lode ergometer to equipment that is not listed hereunder.
Operator manual Corival Cardimax FX-7402 CASE 12, 15, 16 (with Ergometer + BPM) P10 P10 CASE 8000 P10 Cardiosoft P10 Cardiosys P10 MAC1200ST P10 MAC5000 P10 Max I + Corival 400 Max I + Ergometer + BPM P10 P10 Max - PE P4 Heartscreen 112DT LODE Protocol Cardio PC/E LODE Protocol Cardio Perfect LODE Protocol CPX/D System LODE Protocol CPX/MAX/D LODE Protocol CardiO2/CP LODE Protocol Cardio 2/CPX/D LODE Protocol Medgraphics Profiler EX LODE Protocol MedIT AS MeditSense 100 Be
Operator manual Sensormedics Welch Allyn Cardio Control Corival Oxycon Pro Lode WLP Oxycon Alpha Lode WLP Oxycon Beta Lode WLP Cardioperfect 2001 (CP comcard ISA) LODE Protocol Cardioperfect 2001 (Unilink USB or CP comcard PCI) LODE Protocol A Lode ergometer is default shipped with the LODE RS232 communication protocol. You can change the RS232 protocol by following the procedure below: • Switch on the ergometer and press enter.