User`s manual
Chapter 2 Handler Setup
www.exatron.com 3/19/01
2-4
MOD SET UP ? NO
There is a five second time out delay. If no action is taken within 5 seconds the handler will assume no
modification of the setup is necessary and will continue to
READY TO RUN
.
The settings which are accessed by this front panel selection determine certain handler operating
conditions, such as what interface to use in communicating with outside programmers or testers, whether
to stop when a failed device is detected, etc.
[SET UP]
key toggles to
MOD SET UP ? YS
.
[Enter]
key displays:
T SITE = _ _ _ _ _
This setup option sets the number of motor steps from the zero point to the test site. The factory
default is 46585.
Enter changes to the Test Site step count using the controller keypad number keys. The handler
allows changes to each digit from lowest to highest in order. First enter Ones, then Tens, then
Hundreds, etc.
For example:
to enter a step count of 54321
devices, press
[1]
, then
[2]
, then
[3]
, then
[4]
, then
[5]
. The display will read:
T SITE = 54321
.
Press the
[Clear]
key to erase your entry.
Press the
[Enter]
key to set this number into memory and proceed to the next setup option.
FULL TUBE = _ _ _ _
This setup option sets the number of parts for a full output tube. The factory default is 0050.
Enter changes to the Full Tube count using the controller keypad number keys. The handler
allows changes to each digit from lowest to highest in order. First enter Ones, then Tens, then
Hundreds, etc.
For example:
to enter a tube count of 4321
devices, press
[1]
, then
[2]
, then
[3]
, then
[4]
.
The display will read:
FULL TUBE = 4321
.
Press the
[Clear]
key to erase your entry.
Press the
[Enter]
key to set this number into memory and proceed to the next setup option.
PICK INTERFACE ?
[1]
key chooses and displays:
HANDLER PORT
[2]
key chooses and displays:
EXATRON SUPER
[3]
key chooses and displays:
EXATRON RS-232
[4]
key chooses and displays:
PROGRAM RS-232
(Please refer to Interface chapter of the manual for explanation of various interface options. )
[Clear]
key rejects selection and returns to
PICK INTERFACE ?
[Enter]
key accepts selection and proceeds to
STOP ON FAIL ?