User Guide

Calibration
To help ensure accurate glucose readings, you must
routinely calibrate your CGM System with a fingerstick
test from a blood glucose metre. Your CGM System
automatically notifies you when it’s time to calibrate:
24 hours after inserting your sensor, you must
complete 4 fingerstick calibration tests spaced 2 to
12 hours apart.
Every day thereafter you must complete
2 fingerstick calibration tests per day, spaced 10 to
14 hours apart.
Set your daily calibration schedule by tapping
Menu > Settings > Daily Calibration.
How to calibrate:
Do a fingerstick test with a blood glucose
metre. Always follow the metre manufacturer’s
instructions, including washing hands with warm
water, and drying prior to testing.
Enter the fingerstick reading into the app within
10 minutes of testing, ensuring reading and time
are correct.
Alerts
BOTH your mobile device and smart transmitter
provide alerts to notify you when your CGM
readings have reached certain target settings or if
your CGM System requires attention.
Review and confirm alerts on your mobile device.
You can set high/low glucose alerts or target
settings by tapping Menu > Settings >
Glucose.
See the User Guide for a complete listing of alerts
on your app.
Calibrate only when glucose is NOT changing
rapidly (e.g., before meals, before dosing insulin).
The smart transmitter must be worn 5 minutes
before and 15 minutes after the test for a successful
completion of the calibration.