4.2 System Setup
System setup sub-group allows you to configure general settings of the meter. The
settings are displayed in 6 pages. Press NEXT-P (F2) and PREV-P (F1) to navigate
through these pages.
4.2.1 System Settings – Page 1
Figure 68: System Settings - Page 1
Parameter Description Factory Default
STABLE
Indicator
ENABLE - The meter displays ‘Stable’ indicator in the
measurement screen as per the ‘STABLE CRITERIA’
defined below.
DISABLE – ‘Stable’ indicator does not appear.
ENABLE
Stability
Criteria
SLOW – The reading is stabilized slowly and exhibits
good repeatability
MEDIUM – Reading stability is averaged between
slow & fast stability
FAST – Reading is stabilized quickly at the cost of
repeatability.
(This parameter has no effect if ‘STABLE’ parameter
is disabled)
FAST
Auto Hold
ENABLE - The meter holds the reading in the
measurement screen, if the reading is ‘Stable’ for
consecutive 5 seconds.
If this is enabled, ‘Response time’ appears in the
measurement screen, indicating the average response
time of the probe.
DISABLE – The reading is not held
(This parameter has no effect if ‘STABLE’ parameter
is disabled. The response time may not work if the
system time has not been set as described in page
77)
DISABLE