Instruction manual

Instruction Manual PCD 650
73
4.1.2 Accessing Setup mode when password protection enabled
Follow the steps below to access the setup mode, when password protection is
enabled.
1. Switch on the meter. The meter goes to measurement mode.
2. Press right arrow key
to navigate to other functions on the right-side of LCD.
3. Press SETP (F1) to go to Setup mode. Login password screen appears
(Figure 67). The meter expects the 5-digit password specified in system setup.
(Refer in page 80)
Figure 67: Login password screen
4. Press up & down arrow keys to enter the first digit of the password and
then press NEXT (F3) key to move to the next digit.
5. The next digit is selected. Press up & down arrow keys to enter the
second digit of the password. Enter all 5-digits.
6. Press ENTER key to confirm the password.
7. When the correct password is entered, the Setup Key Function Screen
appears.( Figure 65)
8. Press Enter key to launch Setup Selection Screen. (See Figure 66)
9. Press up or down arrow key to go to required setup sub-group.
10. Press ENTER key to select the sub-group.
Note: If you enter an incorrect password, the screen shows ‘Try again’. If an
incorrect password is entered for 3 consecutive times, the meter goes to
measurement mode. If you forget the password there is no way to access the
system setting and calibration.
Note: You can enter ‘00000’ (read-only password) if you wish to view the
setup parameters. You are not allowed to modify any parameter when