User Guide

Instruction Manual PC 10
- 20 -
10. SPECIFICATIONS
Mode pH Temperature Conductivity
Range 0.00 to 14.00 pH 0.0 to 100.0 °C 0 to 19.99 µS
0 to 199.9 µS
0 to 1999 µS
0 to 19.99 mS
Resolution 0.01 pH 0.1 °C 0.01 µS
0.1 µS
1 µS
0.01 mS
Accuracy ± 0.01 pH ± 0.5 °C ±1% Full Scale
+1 digit
Calibration Up to three points
(pH 4.01, 7.00,
10.01) with
automatic buffer
recognition
Offset in 0.1 °C
increments
Up to four points
(one point per
range)
Conductivity Cell
constant (k)
1.0
Conductivity
Temperature
Coefficient
2.00 % per °C
fixed
Temperature
Compensation
Automatic from 0 to 50 °C
Operating
Temperature
0 to 50 °C
Power four 1.5 V AAA-sized batteries (included) or AC adapter
(only for Standard PC 10 meter; optional; order separately on
“Accessories” section)
Battery life >50 hours
Dimensions Waterproof PC 10 meter: 19 cm (L) x 10 cm (W) x 6 cm (H)
Boxed: 24 cm (L) x 23 cm (W) x 7 cm (H)
Probe: 180 mm (L) x 51 mm (Diameter), with 3-m cable
Shipping weight 0.92 kg