User Guide
Instruction Manual CON 400/410
- 48 -
16 ADDENDUM 4: METER FACTORY DEFAULT
SETTINGS
Type Parameter Default Remarks
P1.0 Memory clear No Retains current memory
P2.1 Viewing previous –– No calibration data for 1
st
range
P2.2 calibration data –– No calibration data for 2
nd
range
P2.3 –– No calibration data for 3
rd
range
P2.4 –– No calibration data for 4
th
range
P2.5 –– No calibration data for 5
th
range
P3.1 Viewing probe data 1.0 No offset for effective cell constant (1
st
range)
P3.2 1.0 No offset for effective cell constant (2
nd
range)
P3.3 1.0 No offset for effective cell constant (3
rd
range)
P3.4 1.0 No offset for effective cell constant (4
th
range)
P3.5 1.0 No offset for effective cell constant (5
th
range)
P4.1 Ready indicator / auto
endpoint
READY/ON Ready indicator on; auto endpoint off
P4.2 Select °C/°F °C Degrees Celsius
P4.3 ATC on or off ATC on ––
P4.4 TDS factor 0.50 Adjustable from 0.40 to 1.00 (available in TDS
400 and CON 410 meters only)
P5.1 Temperature coefficient 2.1 % per °
C
Adjustable from 0.0 to 10%
P5.2 Normalisation temperature 25° C Adjustable from 15 to 30° C
P6.1 Cell constant 1.0 Select from k=1.0, 0.1 or 10
P8.0 Factory default No Retains your current settings
NOTES:
The P7.0 clock function retains the year and time that you set, even after reverting to factory
default settings.