User Guide
Instruction Manual CON 400/410
- 43 -
11 SPECIFICATIONS
SPECIFICATIONS DESCRIPTIONS CON 400 CON 410
Conductivity Range
0 to 19.99, 199.9, 1999 µS/cm;
0 to 19.99, 199.9 mS/cm
• •
TDS Range
0.00 to 9.99, 0.0 to 99.9, 0 to 999 ppm;
0.00 to 9.99ppt, 0.0 to 199.9 ppt
•
Resolution 0.05 % Full Scale • •
Accuracy ±1% F.S. ±1 digit • •
0.0 to 100.0 °C
• •
Temperature Range
(32.0 to 212 °F)
0.1 °C
Resolution
(0.1 from 32 °F to 199.9 °F
and 1.0 from 200 °F to 212 °F)
• •
Accuracy
± 0.5 °C (± 0.9 °F)
• •
Cell Constant 0.1, 1.0, 10.0 (selectable) • •
Temperature Compensation Automatic / Manual (from 0 to 80 ° C) • •
Temperature Coefficient 0.0 to 10.0% / ° C • •
Normalization Temperature 15.0 to 30.0 ° C (adjustable) • •
Conductivity to TDS
Conversion factor
0.40 to 1.00
•
Number of calibration points 5: Maximum 1 per range • •
Auto-ranging Yes (manual select available) • •
HOLD Function Yes • •
Memory Yes 50 50
Auto Power off 20 minutes after last key operation • •
Averaging/Stability
(READY)/Auto-hold
Selectable
• •
Inputs 6-pin round connector • •
Display Custom Dual LCD • •
Power Requirements 4 ‘AAA’ Batteries • •
Battery Life > 100 hours • •
Dimension / Weight
Meter: 19 x 10 x 6 cm; 320 g
Boxed: 24 x 23 x 7 cm; 700 g
• •