User Guide

Instruction Manual CON 400/410
- 35 -
7.6.2 P5.2 Adjusting the normalization temperature
Your meter will normalize its conductivity measurements to a standard temperature that you
can select. You can adjust the normalization temperature from 15 to 30 °C (59 to 86 °F). Meter
default is 25 °C (77 °F).
From measurement mode
1. Press SETUP key to enter Set Up mode.
2. Press the MI/S or MR/T keys to scroll through
subgroups until you view parameter P5.0.
3. Press the ENTER key three times to select
parameter 5.2. The display shows “t.nr” on the
upper display.
4. Press the ENTER key again. The upper display
shows the normalisation temperature and the lower
display shows “t.nr”.
5. Press the MI/S or MR/T keys to select the
normalisation temperature.
6. Press the ENTER key to confirm selection and to
return to the subgroup menu. Press CAL/MEAS key
to return to measurement mode.
NOTE:
Most calibration standards are referenced to 25°C. If a
normalization temperature other than 25°C is entered, the
calibration standard value at this temperature must be
known.
Figure 31 - Adjusting the
normalization temperature