User Guide

Instruction Manual CON 400/410
- 32 -
7.5.3 P4.3 Selecting Automatic or Manual Temperature Compensation
This feature lets you select between Automatic Temperature Compensation (ATC) and
Manual Temperature Compensation. Meter default is ATC activated.
From measurement mode
1. Press SETUP key to enter Set Up mode.
2. Press the MI/S or MR/T keys to scroll through
subgroups until you view parameter P4.0.
3. Press the ENTER key three times to select parameter
4.3. The upper display shows “ATC” and the lower
display shows “P4.3”.
4. Press the ENTER key again. The upper display
shows “ATC” and the lower display shows “YES” or
“NO”.
5. Press MI/S or MR/T keys to select Automatic or
Manual Temperature Compensation.
YES = Automatic Temperature Compensation
activated.
NO = Manual Temperature Compensation
activated.
6. Press the ENTER key to confirm selection and to
return to the subgroup menu. Press the CAL/MEAS
key to return to measurement mode.
Figure 28 - Selecting Automatic or
Manual Temperature Compensation