User Guide

Instruction Manual CON 400/410
- 16 -
5.2 Manual Temperature Compensation
IMPORTANT: For manual compensation, you must deactivate the automatic temperature
compensation feature.
5.2.1 Selecting Manual Temperature
Compensation
To select between Automatic Temperature Compensation
(ATC) and Manual Temperature Compensation in the
SETUP program P4.3. Meter default is ATC on.
From measurement mode
1) Press SETUP key to enter Set Up mode.
2) Press the MI/S or MR/T keys to scroll through
subgroups until you view parameter P4.0.
3) Press the ENTER key three times to select parameter
4.3. The upper display shows “ATC” and the lower
display shows “P4.3”.
4) Press the ENTER key again. The upper display shows
“ATC” and the lower display shows “YES” or “NO”.
5) Press the MI/S or MR/T to select between Automatic
or Manual Temperature Compensation feature.
YES = Automatic Temperature Compensation
activated.
NO = Manual Temperature Compensation
activated.
6) Press the ENTER key to confirm selection and to
return to the subgroup menu. Press the CAL/MEAS
key to return to measurement mode.
NOTE:
Manual Temperature Compensation should only be
selected if the ATC fails. For non-compensated
measurements, change the temperature coefficient to
0.0% (see section 7.6.1)
P4.0 Configuration Setup
Selecting Automatic or Manual
Temperature Compensation