Instruction manual

- 30 -
CAL dEF:
Reset all calibration settings of the pH Monitor to
factory defaults, when confirmed by pressing ENT
key.
Press ENT key to confirm your selection.
- If ‘nO dEF’ was selected, the pH Monitor reverts to Cnfg screen.
Press or key to access other setup screens or
press and key simultaneously (escape) to return
to measurement mode.
- If ‘FCt dEF’ or ‘CAL dEF’ was selected, the pH Monitor performs
the selected reset and returns to measurement mode.
Press or key to access other setup screens or press and key
simultaneously (escape) to return to measurement mode.
To exit from any intermediate steps, press and keys simultaneously
(escape). The pH Monitor returns to the first screen configuration settings (Cnfg).
5.6 Viewing Electrode Properties
Each time you calibrate your pH electrode, the pH Monitor re-calculates slope &
offset of the electrode and shown in the LCD at the end of calibration. The setup
mode allows you to view slope & offset values at any time.
NOTE:
When the measurement mode is set to ORP, only the offset value of
the electrode is displayed.
SETUP
HOLD
mV
pH
2
SETUP
ENT
1
ENT
mV
2
SETUP
For pH measurement mode
For ORP measurement mode
ENT