6.4 HOLD Current Setting
When Transmitter is in CAL or SETUP modes, it automatically
goes into a „HOLD‟ mode.
To indicate Transmitter is in „HOLD‟ mode, output current can be
set to 22 mA output by activating the „HLD On‟.
1. Press ENT key and use or keys to scroll till LCD
displays „SEt in the upper display; and HLd in the
lower display. Press ENT again.
2. Upper display now shows HLd. Lower display will
show either OFF or On. Use or keys to toggle
between „On of „OFF. Press ENT to accept selection.
3. Press and keys together, to return to MEAS
mode.
NOTE: If HLd is set to OFF, then current output will be
equivalent to last measured value.
6.5 Out-of-Range Current Setting
If pH or ORP readings exceed the scale, LCD displays Or‟ (over-
range), or Ur (under-range). Transmitter current output can be
set to 3.8 mA to indicate Out-of Range situation.
1. Press ENT key and use or keys to scroll till LCD
displays SEt‟ in the upper display; and org in the lower
display. Press ENT again.
2. Upper display now shows org. Lower display will show
either OFF or On. Use or keys to toggle
between „On of „OFF. Press ENT to accept selection.
3. Press and keys together, to return to MEAS mode.
NOTE: If org is set to OFF, under-range will be set at 4 mA,
while over-range will be set at 20 mA.