2. Press ENT key. Primary display shows ATC‟ and secondary display
shows „On or „OFF. Use ▲ or ▼ keys to select „ATC Off. Press ENT.
3. Primary display shows temperature (default is 25.0), and secondary
display shows P °C‟‟. This is the set up page for setting the temperature of
your transmitter to the temperature of your process or sample by using an
accurate thermometer as reference.
4. Use or keys to set the transmitter to the temperature of your process
or sample. Press ENT key.
5. Primary display shows temperature (default is 25.0), and secondary
display shows C °C‟. This is the set up page to input to your transmitter
the temperature of your calibration solutions.
6. Use ▲ or ▼ keys to set the transmitter to the temperature of your
calibration solutions. Press ENT key.
7. Press and keys together, to return to MEAS mode.
The transmitter will now compensate pH readings for the manually set temperature
(values taken from P °C‟).