Specifications
12
SETUP
HOLD
SETUP
HOLD
pH
READY
pH
6 ADVANCED SETUP FUNCTIONS
The advanced setup mode lets you customize your transmitter‟s preferences and
defaults. This transmitter features different sub groups that organize all setup
parameters.
The sub-groups are:
6.1 Offset Setting (Unavailable in ORP mode)
As a result of the continuous measurement and the application concerned, it may
not be convenient to remove the electrode for calibration. In such cases, an on-line
Offset adjustment is offered.
1. Press ENT key. Transmitter displays „OFS‟. Press
ENT again. Upper display shows amount of
offset, in terms of pH; lower display shows current
measured pH value.
2. Use ▲ or ▼ keys to introduce offset to pH
readings. As the lower measured reading
changes, upper display records the amount of
offset. Maximum of ± 2.00 pH offset is allowed.
3. Press both ▲ or ▼ keys together to return to „
MEAS „ mode.