6 ADVANCED SETUP FUNCTIONS
The advanced setup mode lets you customize your transmitter‟s preferences and
defaults. This transmitter features different sub groups that organize all setup
parameters.
The sub-groups are:
6.1 Offset Setting (Unavailable in ORP mode)
As a result of the continuous measurement and the application concerned, it may
not be convenient to remove the electrode for calibration. In such cases, an on-line
Offset adjustment is offered.
1. Press ENT key. Transmitter displays „OFS. Press
ENT again. Upper display shows amount of
offset, in terms of pH; lower display shows current
measured pH value.
2. Use or keys to introduce offset to pH
readings. As the lower measured reading
changes, upper display records the amount of
offset. Maximum of ± 2.00 pH offset is allowed.
3. Press both or keys together to return to
MEAS „ mode.