Specifications
Instruction Manual Alpha DO 500
13
4. Calibration Mode
4.1 Entering Calibration Mode
While in measurement mode, press CAL key to access calibration mode.
LCD indicates ‘CAL dO’.
Calibration in or
Calibration in saturation
mg/l ppm
%
Calibration Mode
- See Section 4.3
- See Section 4.4
CAL
HOLD
CAL
&
To quit Calibration Mode
ENT
MEAS
%
°C
ATC
READY
NOTES:
x To exit calibration mode at any time during calibration, press and
keys simultaneously (escape). The DO Transmitter returns to the
measurement mode and the old calibration values remain active.
x The calibration is always carried out in the units of measurement (mg/l
or ppm or %) selected in setup mode.