User Guide
12
11 SPECIFICATIONS
Conductivity SPECIFICATIONS pH ORP
FM
CON1
FM
CON2
FM
CON3
Range 0.0 to 14.0 pH 0 to 995 mV 0 to 10.0
mS
0.0 to
990 µS
0.0 to
150 µS
Accuracy ± 0.1 pH ± 5 mV ±0.2
mS/cm
±20
µS/cm
±2
µS/cm
Resolution 0.1 pH 5 mV 0.1
mS/cm
10
µS/cm
1 µS/cm
Calibration Up to 3 points;
pH 4.0, 7.0 & 10.0
Single- point,
Offset ± 100 mV
Single- point,
± 30% full scale
Automatic Buffer Recognition Yes
Automatic Temperature
Compensation
Yes
Operating Temperature 0 to 50 °C (0 to 120 °F)
Inputs BNC BNC Permanently wired
Display 3-digit LED (Red)
Environmental Requirements Operating: 0 to 50 °C (0 to 120 °F)
Storage: - 10 to 60 °C (14 to 140°F)
Humidity Limits: 10 to 95% RH (non condensing)
Power Requirements AC / DC Adapter 9 V, 200 mA
Dimension / Weight Meter: 19 x 10 x 6 cm; 80 g
Boxed: 24 x 23 x 7 cm; 550 g